AquaBplus Data Sheet

Compliance with ISO Standards

The AquaBplus reverse osmosis system streamlines compliance with ISO dialysis water quality standards.

 

ISO 23500-1

Part 1: addresses general requirements for the preparation and quality management of fluids for hemodialysis and related therapies

ISO 23500-2

Part 2: covers water treatment equipment for hemodialysis applications and related therapies.

ISO 23500-3

Part 3: specifies minimum requirements for water used in hemodialysis and related therapies.

The AquaBplus system delivers the operational benefits and technical advantages that renal care facilities demand.

Features

Economical

  • Modular design to fit individual needs

Protective

  • High permeate quality facilitated by a dead-space free tubing connection

Convenient

  • Remote online monitoring available
  • Initial self-testing of all safety-relevant actuators and sensors

Functional

  • Comprehensive data cleansing, quality documentation, and trend monitoring
  • Data logging (service data recording)
  • Detailed error reporting
  • Initial self-testing of all safety-relevant actuators and sensors
  • Internal leakage sensor and integrated leakage monitoring
  • Permeate safety (ring base)
  • Semiautomatic decalcification
  • Ring base for safety
  • Semiautomatic chemical disinfection with HD integration possible
  • Soft membrane start-and-stop capabilities

Options

  • External leakage sensor
  • Infrastructure data management system
  • Fresenius Medical Care Service Software
  • Heat disinfection of up to four dialysis machines in parallel or up to 2.4 L/m of consumption*

Technical Data

Specifications

Hemodialysis Device

Up to 50 machines running at 800 mL/h per device

Permeate Capacity

500–3000 L/h

Efficiency/Yield

Up to 75%

Dimensions (h x w x d)

150 x 55 x 95 cm

Weight (filled)

160–370 kg

Concentrate Pressure

Max. 25 BAR

Permeate Operating Pressure

Max. 6 BAR

Noise Level

Noise level in SUPPLY mode: 62–73 dB (distance of 1 m) (depending on system capacity and features)

Electrical Supply

Electrical Supply/Three-phase Current

208 V 3/N/PE, 60 Hz, 20 A

Power Consumption

AquaBplus 500–1000: 3 kVA

AquaBplus 1500–3000: 4-4.5 kVA

Radiated Heat/Loss

AquaBplus 500–1000: 0.5 kVA

AquaBplus 1500–3000: 0.65 kVA

Overcurrent Protection

20 A tripping characteristic (depending on voltage/version)

D or K or similar recommended (due to high starting currents) Residual current circuit breaker RCD 30 mA recommended

Socket

208 V: hardwired

Type of Protection Against Electric Shock

Protection Class I

Applied Parts Classification

Type B

Degree of Ingress Protection Against Liquids

Drip-proof

Leakage Currents

According to EN 60601-1

Overvoltage Category

II

Pollution Severity

II

Material Group

III b

Operating Mode

Continuous operation (standby)

Product Water Quality

Bacteria (CFU) and Endotoxins (EU)

>99%

Total Dissolved Solids

Product water quality depends on inlet water quality

>96%

Technical Data

Specifications

Feed Pressure

Dynamic 2–6 BAR

Minimum Inlet Flow

Minimum inlet flow in liters per hour at maximum outlet capacity and a yield of 75%

AquaBplus 500: min. 1000 L/h

AquaBplus 1000: min. 2000 L/h

AquaBplus 1500: min. 3000 L/h

AquaBplus 2000: min. 4000 L/h

AquaBplus 2500: min. 5000 L/h

AquaBplus 3000: min. 6000 L/h

Permeate Connection

Direct PE-Xa connector 25 x 3.5 (feed and return) on the system

Inlet Water Connection

1 1/4” external thread, stainlees steel

Drain Water Connection

DN 70 (HT pipe)

Operating Conditions

Water Hardness

<1.0 °dH

Iron

<0.1 mg/L

Manganese

<0.1 mg/L

Chloride

<100 mg/L

Silicate

<25 mg/L

Total Chlorine

0.1 mg/L

Feed Water Conductivity

<2500 uS/cm

Total Salt Content

1500 mg/L

pH

6–8

Silt Density Index

<3

Feed Water Temperature

Min. 5°C/max. 35°C

Atmospheric Pressure

Ambient pressure: 700–1150 hPa

Ambient Temperature Range

+5°C to +35°C

Relative Humidity

Up to 80% at 20°C (non-condensing)

External Connection Options

Ethernet

Electrically isolated interface for data exchange (RJ45) CAT5 The system can be connected to the in-house network

External Start/Stop

Isolated inputs to start the AquaBplus in SUPPLY mode or to stop all operating modes

Volt-free Contacts: Warning, Alarm, Supply

24V AC/DC/1A

External Failure

Electrically isolated input as “collective alarm” from ext. equipment

External Leakage

Any additional equipment connected to this system must comply with the applicable IEC or ISO standards (such as IEC 60950-1 for information technology equipment). Plus, all system configurations must comply with the requirements for medical systems (according to Annex I to EN 60601-1).

e.g., AquaDETECTOR

Technical Data

Transport and Storage Conditions

Storage Temperature Range

+5°C to +40°C (protect from freezing)

Storage Time

Storage time of preserved system: maximum 12 months

Atmospheric Pressure

Ambient pressure: 500–1150 hPa

Relative Humidity

Up to 80% at 20°C (non-condensing)

Materials in Contact with Dialysis Water

Type

 

Material

Polymers

PP

Polypropylene Polyethylene Polysulfone Polyphenylene oxide

Polyvinylidene fluoride

 

PE

 

PSU

 

PPO

 

PVDF

Rubber

EPDM

Ethylene propylene diene monomer Silicone

Metals

1.4571

Stainless steel Stainless steel

Titanium

 

1.4404

 

Ti

Ceramics

AI2O3

Ceramic

Indications for Use

The AquaBplus Water Purification Systems are reverse osmosis units intended for use with hemodialysis systems to remove organic and inorganic substances and microbial contaminants from the water used for treating hemodialysis patients or other related therapies. These devices are intended to be a component in a complete water purification system, and are not complete water treatment systems. Each reverse osmosis unit must be preceded by pre-treatment devices, and may need to be followed by post-treatment devices as well, to meet current AAMI/ANSI/ISO and federal (U.S.) standards.